DocumentCode :
679116
Title :
Digital data processing of stilbene
Author :
Amiri, Mozhgan ; Cvachovec, Frantisek ; Matej, Zdenek ; Mravec, Filip ; Prenosil, Vaclav
Author_Institution :
Fac. of Inf., Masaryk Univ., Brno, Czech Republic
fYear :
2013
fDate :
23-27 June 2013
Firstpage :
1
Lastpage :
7
Abstract :
Stilbene is a proven spectrometric detector for mixed fields of neutrons and gamma rays. By digital processing of shape output pulses from the detector it is possible to obtain information about the energy of the interacting neutron / photon and distinguish which of these two particles interacts in the detector. Another numerical processing of digital data can finalize the energy spectrum of both components of the mixed field. The quality of the digitized data is highly dependent on the parameters of the hardware used for digitization and on the quality of software processing. Our results also show how the quality of the particle type identification depends on the sampling rate and as well as the method of processing of the sampled data.
Keywords :
analogue-digital conversion; solid scintillation detectors; digital data processing; energy spectrum; numerical processing; particle-type identification; sampling rate; software processing; spectrometric detector; stilbene organic scintillator; Anodes; Detectors; Image edge detection; Neutrons; Photonics; Resistance; Shape; Digitization; digitizer resolution; filtering; neutron/gamma discrimination; pulse length; sampling rate; stilbene; two-parameter spectrometer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), 2013 3rd International Conference on
Conference_Location :
Marseille
Print_ISBN :
978-1-4799-1046-5
Type :
conf
DOI :
10.1109/ANIMMA.2013.6728010
Filename :
6728010
Link To Document :
بازگشت