Abstract :
This paper presents a passive-marker-based optical tracking system utilizing dual Kinect sensors and additional custom optical tracking components. To obtain sub-millimeter tracking accuracy, we introduce robust calibration of dual infrared sensors and point correspondence establishment in a stereo configuration. The 3D localization is subsequently accomplished using multiple back projection lines. The proposed system extends existing inexpensive consumer electronic devices, implements tracking algorithms, and shows the feasibility of applying the proposed low-cost system to surgical training for computer assisted surgeries.
Keywords :
biomedical equipment; calibration; infrared detectors; optical tracking; surgery; calibration; computer assisted surgeries; dual infrared sensors; dual kinect sensors; marker-based surgical instrument tracking; multiple back projection lines; passive-marker-based optical tracking system; submillimeter tracking accuracy; Accuracy; Calibration; Cameras; Instruments; Sensors; Surgery; Three-dimensional displays; Kinect sensor; navigated surgery; optical tracking; tracking;
fLanguage :
English
Journal_Title :
Automation Science and Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
1545-5955
Type :
jour
DOI :
10.1109/TASE.2013.2283775
Filename :
6648452
Link To Document :
بازگشت