Measurement of energy and timing resolution of very highly pixellated LYSO crystal blocks with multiplexed SiPM readout for use in a small animal PET/MR insert
Author :
Thompson, C.J. ; Goertzen, A.L. ; Kozlowski, P. ; Retiere, F. ; Stortz, G. ; Sossi, V. ; Xuezhu Zhang
Author_Institution :
Montreal Neurological Inst., McGill Univ., Montreal, QC, Canada
fYear :
2013
fDate :
Oct. 27 2013-Nov. 2 2013
Firstpage :
1
Lastpage :
5
Abstract :
Arrays of silicon photo-multipliers (SiPMs) are good candidates for the readout of detectors in PET MR inserts due to their high packing density, efficiency, low bias voltage and insensitivity to magnetic fields. In this study we report the readout performance of SensL SiPM arrays in terms of their ability to resolve all elements of pixellated lutetium oxy-orthosilicate (LYSO) crystals, and their energy and timing resolution. A SensL SB4-300-35-CER SiPM array consisting of sixteen 3 × 3 mm elements were used as light sensor. An LYSO crystal block consisting of 10×10 1.2 mm × 1.2 mm × 6.0 mm crystals on the bottom layer and 9×9 1.2 mm × 1.2 mm × 4.0 mm crystals on the top layer (which is offset by ½ the crystal width) was mounted on the SensL array, and covers over 95% of its area. A DPC encoding multiplexor with HDMI cable readout made by the Triumf Instrumentation group with a footprint suitable for allowing 16 modules to form a ring inside a Brooker 7T MR rodent imager was used as a readout. The HDMI cable supplies power, and provides readout of four channels and the device temperature. The average energy resolution for all 100 crystals in the lower layer was 11.3±1.8% and 10.8±1.2% for the 81 crystals in the upper layer. The average timing resolution for all 100 crystals in the lower layer was 2.52±0.23 nsec. and 2.55±0.23 nsec. for the 81 crystals in the upper layer.
Keywords :
biomedical MRI; biomedical equipment; photodetectors; photomultipliers; positron emission tomography; silicon radiation detectors; DPC encoding multiplexor; HDMI cable readout; HDMI cable supplies power; SensL SB4-300-35-CER SiPM array; Triumf Instrumentation group; animal PET-MR insert; energy measurement; energy resolution; high-packing density; high-packing efficiency; light sensor; low bias voltage; magnetic fields; magnetic flux density 7 T; multiplexed SiPM readout; pixellated lutetium oxyorthosilicate crystals; silicon photomultiplier arrays; timing resolution; very-highly pixellated LYSO crystal blocks; Crystals; Detectors; Energy resolution; Image resolution; Magnetic resonance imaging; Positron emission tomography; Timing; PET; Silicon Photo-multiplier; Small Animal PET insert for MRI scanner;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location :
Seoul
Print_ISBN :
978-1-4799-0533-1
Type :
conf
DOI :
10.1109/NSSMIC.2013.6829134
Filename :
6829134
Link To Document :
بازگشت