DocumentCode
686707
Title
One-pair prototype integrated system of DOI- PET and the RF-coil specialized for simultaneous PET-MRI measurements
Author
Nishikido, Fumihiko ; Obata, T. ; Inadama, Naoko ; Yoshida, Erika ; Suga, Mikio ; Shimizu, Kazuo ; Tachibana, Ayako ; Ito, H. ; Yamaya, Taiga
Author_Institution
Nat. Inst. of Radiol. Sci., Chiba, Japan
fYear
2013
fDate
Oct. 27 2013-Nov. 2 2013
Firstpage
1
Lastpage
3
Abstract
We are developing a PET system integrated with a birdcage RF coil for PET-MRI to realize highly sensitive PET system for simultaneous measurements. The PET detectors which consist of a scintillator block, photo sensors and front-end circuits with four-layer DOI encoding capability are placed close to the objective. Therefore, the proposed system can achieve high sensitivity without degradation spatial resolution at the edge of the field of view due to parallax error thanks to the four-layer DOI capability. If a shielding material is inside the RF coils, the RF pulse is blocked by the shielding material and then complete images cannot be obtained. Therefore, each RF coil element is inserted between the scintillator crystal blocks and then medially located in the shielding materials. We developed an one-pair prototype system for performance evaluation of PET detector for PET-MRI.
Keywords
biomedical MRI; coils; optical sensors; positron emission tomography; scintillation counters; DOI-PET system; PET-MRI measurements; birdcage RF coil; front-end circuits; magnetic resonance imaging; one-pair prototype integrated system; parallax error; photo sensors; positron emission tomography; scintillator crystal blocks; Arrays; Crystals; Detectors; Magnetic resonance imaging; Positron emission tomography; Prototypes; Radio frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location
Seoul
Print_ISBN
978-1-4799-0533-1
Type
conf
DOI
10.1109/NSSMIC.2013.6829136
Filename
6829136