• DocumentCode
    686740
  • Title

    Simulation study of the DOI-PEM scanner

  • Author

    Kobayashi, S. ; Yamamoto, Takayuki ; Kajino, Fumiyoshi

  • Author_Institution
    Dept. of Phys., Saga Univ., Saga, Japan
  • fYear
    2013
  • fDate
    Oct. 27 2013-Nov. 2 2013
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A simple MC simulation using the Mathematica app. is presented for a square-shaped PEM device consisted of the notched scintillating crystal chips. For a previously studied material of a 30 mm length Pr:LuAG with the cross sectional area of either 2×2 mm2 or 4×4 mm2, the imaging spread and the detection efficiency of a point source are examined. Chosen the number of DOI notches as a parameter, an optimum solution to give the sub-millimeter space resolution is obtained. Some more results for the PEM construction also are presented.
  • Keywords
    Monte Carlo methods; image resolution; lutetium compounds; mammography; praseodymium; solid scintillation detectors; DOI notches; DOI-PEM scanner; LuAG:Pr; Mathematica app; PEM construction; cross sectional area; detection efficiency; imaging spread; notched scintillating crystal chips; optimum solution; point source; simple MC simulation; simulation study; size 30 mm; square-shaped PEM device; submillimeter space resolution; Apertures; Crystals; Image edge detection; Image resolution; Image segmentation; Imaging; Mathematical model;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
  • Conference_Location
    Seoul
  • Print_ISBN
    978-1-4799-0533-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2013.6829169
  • Filename
    6829169