• DocumentCode
    686775
  • Title

    Low intensity fluorescence light measurements using Silicon Photomultiplier with dedicated front-end ASIC

  • Author

    Baszczyk, Mateusz ; Dorosz, Piotr ; Glab, Sebastian ; Mik, L. ; Kucewicz, Wojciech ; Pijanowska, D.G. ; Szczypinski, R.

  • Author_Institution
    Dept. of Electron., AGH Univ. of Sci. & Technol., Krakow, Poland
  • fYear
    2013
  • fDate
    Oct. 27 2013-Nov. 2 2013
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    The paper presents front-end ASIC and measurements of low intensity fluorescence light using Silicon Photomultiplier. Front-end ASIC is dedicated device for amplifying and shaping signals from Silicon Photomultipliers. Measurement method is described. Fluorescence intensity for sodium fluoresceinate and resorufin in different concentrations is presented. Sensitivity limit is studied.
  • Keywords
    application specific integrated circuits; biomedical equipment; biomedical measurement; dyes; fluorescence; medical signal processing; optical variables measurement; photomultipliers; sensitivity; silicon; sodium compounds; Si; dedicated front-end ASIC; low-intensity fluorescence light measurements; resorufin concentrations; sensitivity limit; signal amplification; signal shaping; silicon photomultiplier; sodium fluoresceinate concentrations; Application specific integrated circuits; Biomedical measurement; Fluorescence; Optical filters; Photomultipliers; Photonics; Silicon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
  • Conference_Location
    Seoul
  • Print_ISBN
    978-1-4799-0533-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2013.6829205
  • Filename
    6829205