Title :
Evaluation of the novel 3D SPECT modelling algorithm in the STIR reconstruction framework: Simple vs. full attenuation correction
Author :
Marti-Fuster, Berta ; Erlandsson, Kjell ; Falcon, Carles ; Tsoumpas, Charalampos ; Livieratos, Lefteris ; Ros, David ; Thielemans, Kris
Author_Institution :
Physiol. Sci. Dept. I, Univ. of Barcelona, Barcelona, Spain
fDate :
Oct. 27 2013-Nov. 2 2013
Abstract :
A 3D SPECT system matrix modelling library has recently been incorporated into STIR - a software package for tomographic image reconstruction. The SPECT modelling accounts for the effects of attenuation and spatially variant resolution by incorporating them into the system matrix. Attenuation calculation can be done either along a single line (simple model), or along multiple lines for the various detector pixels covered by the PSF (full model). Due to practical reasons, a simple central-line approximation is often used for the attenuation modelling in SPECT. Our aim was to evaluate the effect of this approximation in the reconstructed images using STIR. A rotating SPECT system was modelled, equipped with different collimators. We generated noise-free projection data using the full model and reconstructed images using both models. We used a phantom consisting of an ellipse containing 4 spherical inserts with the same activity concentration but different attenuation coefficients. Images were reconstructed using the OS-MAP algorithm, 12 subsets and up to 100 iterations. With the simple model, the four spheres were all different in terms of intensity as well as distribution. When using the full attenuation model, all spheres appeared quite similar, independent of the attenuation. Our results show that the simple attenuation model can lead to artifacts and inaccurate quantification, while the full model adds significant accuracy and stability to the reconstructed images.
Keywords :
approximation theory; collimators; image reconstruction; maximum likelihood estimation; medical image processing; phantoms; physiological models; single photon emission computed tomography; 3D SPECT system matrix modelling library; OS-MAP algorithm; PSF; STIR reconstruction framework; activity concentration; artifacts; attenuation calculation; attenuation coefficients; attenuation modelling; collimators; detector pixels; full attenuation correction; full attenuation model; multiple lines; noise-free projection data; phantom; reconstructed image accuracy; reconstructed image stability; rotating SPECT system; simple attenuation correction; simple central-line approximation; single line; software package for tomographic image reconstruction; spatially variant resolution; spherical inserts; Attenuation; Collimators; Detectors; Image reconstruction; Image resolution; Object oriented modeling; Single photon emission computed tomography;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location :
Seoul
Print_ISBN :
978-1-4799-0533-1
DOI :
10.1109/NSSMIC.2013.6829258