Abstract
The increase on the collective radiation dose due to the large number of medical imaging exams has led the medical physics community to deeply consider the amount of dose delivered to patients as well as its associated risks in these exams. For this purpose we have developed a Monte Carlo tool, PENRADIO, based on a modified version of the 2006 release of the PENELOPE code, to obtain an accurate individualized radiation dose in conventional and interventional radiography and in computed tomography (CT). This tool has been validated showing excellent agreement between the measured and simulated organ doses in the case of a hip conventional radiography and a coronography. We expect the same accuracy in further results for other localizations and in CT examinations.
Keywords
Monte Carlo methods; computerised tomography; dosimetry; Monte Carlo PENRADIO software; PENELOPE code; computed tomography; coronography; dose calculation; hip conventional radiography; interventional radiography; medical imaging; Aluminum; Biomedical imaging; Computed tomography; Electron tubes; Geometry; Monte Carlo methods; Software;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location
Seoul
Print_ISBN
978-1-4799-0533-1
Type
conf
DOI
10.1109/NSSMIC.2013.6829369
Filename
6829369
Link To Document