DocumentCode :
686977
Title :
Timing and intermediate strip studies of a high spatial resolution Resistive Plate Chamber
Author :
Cussans, D. ; Baesso, P. ; Thomay, C. ; Velthuis, J.J. ; Burns, Jack ; Quillin, S. ; Stapleton, M. ; Steer, C.
Author_Institution :
Univ. of Bristol, Bristol, UK
fYear :
2013
fDate :
Oct. 27 2013-Nov. 2 2013
Firstpage :
1
Lastpage :
3
Abstract :
Resistive Plate Chambers(RPCs) offer simple construction and efficient detection of charge particles over a large area with nano-second precision. Traditionally, large strips or pads that cover the entire area of induced charge are used to pick up the signal. The signal is then connected to a binary read-out. This approach gives moderate spatial resolution, typically of the order of a centimetre. However using fine-pitch pick-up strips, with the induced signal spread over several strips and analogue read-out, allows much better spatial resolution to be obtained than is possible with wide strips and binary readout.
Keywords :
fine-pitch technology; nuclear electronics; position sensitive particle detectors; readout electronics; analogue readout; binary readout; charge particle detection; fine-pitch pick-up strips; high spatial resolution resistive plate chamber; induced charge; induced signal; nanosecond precision; Current measurement; Strips; Switches;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location :
Seoul
Print_ISBN :
978-1-4799-0533-1
Type :
conf
DOI :
10.1109/NSSMIC.2013.6829414
Filename :
6829414
Link To Document :