DocumentCode :
687152
Title :
Performance of X-ray detectors with SiPM readout in cargo accelerator-based inspection systems
Author :
Vinogradov, Sergey ; Arodzero, Anatoli ; Lanza, Richard C.
Author_Institution :
Dept. of Phys., Univ. of Liverpool, Warrington, UK
fYear :
2013
fDate :
Oct. 27 2013-Nov. 2 2013
Firstpage :
1
Lastpage :
6
Abstract :
Existing requirements for high throughput cargo radiography inspection include high resolution images (better than 5 mm line pair resolution), penetration beyond 400 mm steel equivalent, material discrimination (organic, inorganic, high Z), high scan speeds (> 10kph, up to 60kph), low dose and small radiation exclusion zone; all in a cost effective system. To meet and exceed these requirements research into a number of new radiography methods has been initiated. Novel concepts relying on intrapulse modulated energy X-ray sources, mono-energetic gamma-ray sources, and new types of fast X-ray detectors, Scintillation-Cherenkov Detectors, are expected to be more beneficial being combined with unique features of Silicon Photomultiplier (SiPM) technology.
Keywords :
X-ray detection; elemental semiconductors; inspection; photomultipliers; silicon; Si; SiPM readout; X-ray detectors; afterpulsing; cargo accelerator-based inspection systems; crosstalk; high dark noise; photon detector; pixel recovery transients; scintillator; silicon photomultiplier; Analytical models; Detectors; Dynamic range; Inspection; Linearity; Noise; Photonics;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location :
Seoul
Print_ISBN :
978-1-4799-0533-1
Type :
conf
DOI :
10.1109/NSSMIC.2013.6829597
Filename :
6829597
Link To Document :