• DocumentCode
    694545
  • Title

    A calibration System for artificial ear in telecommunication

  • Author

    Hu Bingxin ; Zhu Haijiang ; Zhong Bo ; Yang Ping ; He Longbiao ; Xu Huan

  • Author_Institution
    Coll. of Inf. Sci. & Technol., Beijing Univ. of Chem. & Technol., Beijing, China
  • fYear
    2013
  • fDate
    12-13 Oct. 2013
  • Firstpage
    1128
  • Lastpage
    1132
  • Abstract
    In this paper we propose a novel system for automatic calibration for the artificial ears. Aiming at the artificial ears for telephonometric use, the system is designed in accordance with the international standards. The calibration process involves testing the parameters of the probe microphone and the acoustic characteristics of the artificial ear on open and closed ear condition. The parameters include sound pressure level, acoustic impedance, frequency response sensitivity, etc. On above basis, all processes are controlled by the controlling program. The measured data can be analyzed automatically by this program. As a result, the final results can be obtained. The results show that the parameters measured for 318 ears conform to the recommendation ITU-T P.57. Experiments indicate that this system can accurately realize the artificial ear calibration.
  • Keywords
    acoustic signal processing; calibration; ear; ITU-T P.57; acoustic characteristics; acoustic impedance; artificial ears; automatic calibration; frequency response sensitivity; probe microphone; sound pressure level; telephonometric use; Acoustic measurements; Calibration; Ear; Frequency measurement; Microphones; Probes; Sensitivity; artificial ear; automatic calibration; microphone;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computer Science and Network Technology (ICCSNT), 2013 3rd International Conference on
  • Conference_Location
    Dalian
  • Type

    conf

  • DOI
    10.1109/ICCSNT.2013.6967301
  • Filename
    6967301