DocumentCode :
69883
Title :
Magnetically Inserted Neural Electrodes: Tissue Response and Functional Lifetime
Author :
Dryg, Ian D. ; Ward, Matthew P. ; Qing, Kurt Y. ; Mei, Henry ; Schaffer, Jeremy E. ; Irazoqui, Pedro P.
Author_Institution :
Dept. of Bioeng., Univ. of Washington, Seattle, WA, USA
Volume :
23
Issue :
4
fYear :
2015
fDate :
Jul-15
Firstpage :
562
Lastpage :
571
Abstract :
Neural recording and stimulation have great clinical potential. Long-term neural recording remains a challenge, however, as implantable electrodes eventually fail due to the adverse effects of the host tissue response to the indwelling implant. Astrocytes and microglia attempt to engulf the electrode, increasing the electrical impedance between the electrode and neurons, and possibly pushing neurons away from the recording site. Faster insertion speed, finer tip geometry, smaller size, and lower material stiffness all seem to decrease damage caused by insertion and reduce the intensity of the tissue response. However, electrodes that are too small result in buckling, making insertion impossible. In this paper, we assess the viability of high-speed (27.8 m/s) deployment of 25 μm, ferromagnetic microelectrodes into rat brain. To characterize functionality of magnetically inserted electrodes, 4 Long-Evans rats were implanted for 31 days with impedance measurements and neural recordings taken daily. Performance was compared to 150 μm diameter PlasticsOne electrodes since 25 μm electrodes buckled during “slow speed” insertion. Platinum-iron magnetically inserted electrodes resolved single unit activity throughout the duration of the study in one rat, and saw no significant change (p=0.970) in impedance (4.54% increase) from day 0 (Z0 ≈ 144 kΩ,Z31 ≈ 150 kΩ). These findings provide a proof-of-concept for magnetic insertion as a viable insertion method that enables nonbuckling implantation of small (25 μm) microelectrodes, with potential for neural recording applications.
Keywords :
bioelectric phenomena; biological tissues; biomedical electrodes; brain; electric impedance measurement; iron alloys; microelectrodes; neurophysiology; platinum alloys; Long-Evans rats; PlasticsOne electrodes; PtFe; astrocytes; electrical impedance; ferromagnetic microelectrodes; functional lifetime; host tissue response; impedance measurements; implantable electrodes; magnetically inserted neural electrodes; microglia; neural recording; neural stimulation; nonbuckling implantation; rat brain; single unit activity; size 150 mum; size 25 mum; time 31 day; tissue response; Brain; Electrodes; Electron tubes; Impedance; Magnetic recording; Rats; Electrode; impedance; magnetic insertion; neural recording; tissue response;
fLanguage :
English
Journal_Title :
Neural Systems and Rehabilitation Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
1534-4320
Type :
jour
DOI :
10.1109/TNSRE.2015.2399856
Filename :
7044555
Link To Document :
بازگشت