• DocumentCode
    714078
  • Title

    Class E RF amplifier design in an ultrasonic link for wireless power delivery to implanted medical devices

  • Author

    Vihvelin, Hugo ; Leadbetter, Jeff ; Bance, Manohar ; Brown, Jeremy A. ; Adamson, Robert B. A.

  • Author_Institution
    Sch. of Biomed. Eng., Dalhousie Univ., Halifax, NS, Canada
  • fYear
    2015
  • fDate
    3-6 May 2015
  • Firstpage
    382
  • Lastpage
    386
  • Abstract
    Ultrasonic power transfer is a promising alternative to electromagnetic induction for providing wireless power to active implanted medical devices. In a portable ultrasonic power link for devices like cochlear implants, there is a key requirement for an inverter circuit that is capable of driving the transmitting piezoelectric transducer with high efficiency in order to maximize battery life and overall device reliability. In the following paper, a high efficiency Class E amplifier is presented for use on the transmit side of an ultrasonic transcutaneous energy transfer (UTET) link. The amplifier is characterized by a peak efficiency of 90.8% at an output power level of 293 mW and drive frequency of 1.275 MHz.
  • Keywords
    biomedical equipment; invertors; piezoelectric transducers; power supplies to apparatus; radiofrequency amplifiers; ultrasonic transducers; UTET link; class E RF amplifier design; electromagnetic induction; implanted medical device; inverter circuit; piezoelectric transducer; ultrasonic power link; ultrasonic power transfer; ultrasonic transcutaneous energy transfer link; wireless power delivery; Acoustics; Inverters; Logic gates; Power amplifiers; Power generation; Resonant frequency; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electrical and Computer Engineering (CCECE), 2015 IEEE 28th Canadian Conference on
  • Conference_Location
    Halifax, NS
  • ISSN
    0840-7789
  • Print_ISBN
    978-1-4799-5827-6
  • Type

    conf

  • DOI
    10.1109/CCECE.2015.7129307
  • Filename
    7129307