Metrological Performances of a Plantar Pressure Measurement System
Author
Saggin, Bortolino ; Scaccabarozzi, Diego ; Tarabini, Marco
Author_Institution
Politec. di Milano, Milan, Italy
Volume
62
Issue
4
fYear
2013
fDate
Apr-13
Firstpage
766
Lastpage
776
Abstract
Plantar pressure measurements provide useful information to diagnose a diverse range of foot disorders; unfortunately, the commercially available measurement systems are undesirably sensitive to several disturbances, but this aspect is mostly neglected in the literature. This paper describes the results of an experimental campaign aiming at the identification of pressure measuring system metrological performances, at system modeling, and at the implementation of correction procedures. The sensor model was implemented using the results of static and dynamic tests performed on a pedar-X plantar pressure measurement system. The static calibration was performed by analyzing the effect of temperature, single sensor coverage area, local curvature, tangential forces, long-term stability (creep), and sensor crosstalk on the system performances. The dynamic calibration was performed on an electrodynamic shaker, identifying the single sensor frequency response function and the hysteresis under different average loads. The dynamic sensor model is based on the Kelvin-Voigt model, which is representative of the viscoelastic behavior of the material. The model allowed us to compensate both the creep (i.e., the behavior under static loads) and the nonunitary frequency response function. A deconvolution-based algorithm has been proposed to compensate the sensor crosstalk effects, although its implementation requires additional investigations. Experimental results of bobbing and gait tests showed that, with the adoption of the proposed compensation algorithms, the force and center of pressure errors could be reduced by more than 50% of their initial values.
Keywords
electrodynamics; pressure measurement; Kelvin-Voigt model; correction procedures; dynamic sensor model; dynamic tests; effect of temperature; electrodynamic shaker; foot disorders; metrological performances; pedar-X plantar pressure measurement system; single sensor coverage area; static calibration; static tests; tangential forces; Biomedical measurements; error compensation; gait analysis; plantar pressure; uncertainty;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2013.2245185
Filename
6471220
Link To Document