DocumentCode :
716968
Title :
Numerical analysis of EMF safety and thermal aspects in a pacemaker with a Wireless Power Transfer system
Author :
Campi, Tommaso ; Cruciani, Silvano ; Santilli, Gilda Paolini ; Feliziani, Mauro
Author_Institution :
Dept. of Ind. & Inf. Eng. & Econ., Univ. of L´Aquila, L´Aquila, Italy
fYear :
2015
fDate :
13-15 May 2015
Firstpage :
1
Lastpage :
4
Abstract :
This study deals with the numerical dosimetry analysis of a Wireless Power Transfer (WPT) system based on magnetic resonant coupling applied to a pacemaker for recharge its battery. Considering a quite low operational frequency (f=300 kHz), it is possible to obtain good results in terms of transferred power within a low impact on EMC and EMF safety issues. The numerical study permits to evaluate the temperature increase in the human body model by a thermal analysis.
Keywords :
dosimetry; electric potential; electromagnetic compatibility; inductive power transmission; numerical analysis; pacemakers; thermal analysis; EMC; EMF safety; WPT system; frequency 300 kHz; magnetic resonant coupling; numerical dosimetry analysis; pacemaker; wireless power transfer system; Biological system modeling; Biological tissues; Magnetic fields; Magnetic noise; Magnetic shielding; Pacemakers; Wireless communication; EMF safety; SAR; magnetic resonant coupling; pacemaker; temperature increase; wireless power transfer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Wireless Power Transfer Conference (WPTC), 2015 IEEE
Conference_Location :
Boulder, CO
Type :
conf
DOI :
10.1109/WPT.2015.7140172
Filename :
7140172
Link To Document :
بازگشت