DocumentCode :
717383
Title :
Focused ultrasound temperature effect in tissue-mimicking material and sheep liver
Author :
Karaboce, Baki
Author_Institution :
TUBITAK, Nat. Metrol. Inst., Kocaeli, Turkey
fYear :
2015
fDate :
7-9 May 2015
Firstpage :
131
Lastpage :
134
Abstract :
The temperature distribution produced by a HIFU (High Intensity Focused Ultrasound) transducer has been investigated by a set up using T type thermocouple temperature sensors in the automated scanning system. The temperature change and the temperature distribution due to applied ultrasound power have been characterized. The temperature measurements were carried out in a tissue-mimicking material (TMM) that has characteristics very similar to human tissue and in ex vivo tissue. A temperature measurement set up in a tissue phantom has been constructed at TÜBİTAK UME (The Scientific and Technological Research Council of Turkey, the National Metrology Institute) Ultrasound laboratory. Burning and removing the tissue by high power ultrasound have been showed.
Keywords :
biological tissues; biomedical transducers; biomedical ultrasonics; biomimetics; biothermics; liver; phantoms; temperature distribution; temperature measurement; temperature sensors; thermocouples; ultrasonic therapy; HIFU transducer; T type thermocouple temperature sensors; automated scanning system; focused ultrasound temperature distribution; high intensity focused ultrasound; sheep liver; temperature measurements; tissue phantom; tissue-mimicking material; Phantoms; Temperature distribution; Temperature measurement; Temperature sensors; Transducers; Ultrasonic imaging; Ultrasonic variables measurement; HIFU; TMM; metrology; temperature measurements; tissue phantom;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Medical Measurements and Applications (MeMeA), 2015 IEEE International Symposium on
Conference_Location :
Turin
Type :
conf
DOI :
10.1109/MeMeA.2015.7145186
Filename :
7145186