    iNODE in-vivo testing for selective vagus nerve recording and stimulation

  • Author

    Cota, Oscar F. ; Schlosser, Mario ; Schiek, Michael ; Stieglitz, Thomas ; Gierthmuehlen, Mortimer ; Plachta, Dennis

  • Author_Institution
    Dept. of Microsyst. Eng., Univ. of Freiburg, Freiburg, Germany
  • fYear
    2015
  • fDate
    22-24 April 2015
  • Firstpage
    525
  • Lastpage
    528
  • Abstract
    Hypertension is a wide spread disease and despite pharmacological treatment some patients cannot obtain any relief from medication. In previous work, we could show that blood pressure can be reduced by Vagus Nerve Stimulation (VNS) almost free from side effects. In this work, we focus on the implementation of a miniaturized system using a modified embedded sensor/actuator device iNODE for vagus nerve recording and stimulation. The results illustrate that this device, using a 3.3 V power supply, 58.9 mW while stimulating, 7.8 mW while recording, is capable of reducing the blood pressure, showing equivalent results to major laboratory equipment.
  • Keywords
    biomedical electrodes; medical disorders; medical signal processing; neurophysiology; patient treatment; blood pressure; disease; hypertension; iNODE in-vivo testing; modified embedded sensor-actuator device; pharmacological treatment; side effects; vagus nerve recording-stimulation; Conferences; Neural engineering;