Title :
Dependence of destruction from the characteristics of radiofrequency impact and length of the electrode
Author :
Sychyk, M.M. ; Maksymenko, V.B. ; Perepeka, E.O. ; Kravchuk, B.B. ; Batsak, B.V.
Author_Institution :
Fac. of Biomed. Eng., Nat. Tech. Univ. of Ukraine, Kiev, Ukraine
Abstract :
The work is devoted to research of catheter radiofrequency ablation modes (CRFA) in the treatment of the heart pathways disease. Lesions sizes of myocardial tissue during CRFA have been obtained in the experiment depending on output of the generator power, the electrode temperature, the size of electrode tip, duration of RF exposure. The work gives suggestions for optimal catheter radiofrequency ablation modes.
Keywords :
biothermics; cardiology; catheters; diseases; muscle; surgery; CRFA; RF exposure duration; catheter radiofrequency ablation modes; destruction; electrode length; electrode temperature; electrode tip size; generator power; heart pathways disease treatment; lesions sizes; myocardial tissue; optimal catheter radiofrequency ablation modes; radiofrequency impact; Catheters; Cooling; Electrodes; Geometry; Lesions; Myocardium; Radio frequency; catheter radiofrequency ablation; electrode temperature; generator power; lesions sizes; myocardial tissue; optimal ablation modes;