DocumentCode
720098
Title
Power meter verification issue: Reactive power measurement in non sinusoidal conditions
Author
Gallo, Daniele ; Landi, Carmine ; Luiso, Mario
Author_Institution
Dept. of Ind. & Inf. Eng., Second Univ. of Naples, Aversa, Italy
fYear
2015
fDate
11-14 May 2015
Firstpage
1255
Lastpage
1260
Abstract
In this paper, the authors propose a methodology to guide verification of single phase reactive power meters under non sinusoidal conditions. At first, it is presented a brief review about different definitions of reactive power proposed in literature. Then a test signal for the detection of the implemented approach without a priori knowledge provided by the manufacturer was designed and numerically tested. Then the proposed methodology was applied to some commercial power meter showing preliminary results.
Keywords
numerical analysis; power measurement; power meters; reactive power; nonsinusoidal conditions; numerical testing; reactive power measurement; single phase reactive power meter guide verification method; Accuracy; Current measurement; Harmonic analysis; Power measurement; Power system harmonics; Reactive power; Voltage measurement; energy meters verification; measurement; non-sinusoidal conditions; quality assurance;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference (I2MTC), 2015 IEEE International
Conference_Location
Pisa
Type
conf
DOI
10.1109/I2MTC.2015.7151453