  • Title

    Power meter verification issue: Reactive power measurement in non sinusoidal conditions

  • Author

    Gallo, Daniele ; Landi, Carmine ; Luiso, Mario

  • Author_Institution
    Dept. of Ind. & Inf. Eng., Second Univ. of Naples, Aversa, Italy
  • fYear
    2015
  • fDate
    11-14 May 2015
  • Firstpage
    1255
  • Lastpage
    1260
  • Abstract
    In this paper, the authors propose a methodology to guide verification of single phase reactive power meters under non sinusoidal conditions. At first, it is presented a brief review about different definitions of reactive power proposed in literature. Then a test signal for the detection of the implemented approach without a priori knowledge provided by the manufacturer was designed and numerically tested. Then the proposed methodology was applied to some commercial power meter showing preliminary results.
  • Keywords
    numerical analysis; power measurement; power meters; reactive power; nonsinusoidal conditions; numerical testing; reactive power measurement; single phase reactive power meter guide verification method; Accuracy; Current measurement; Harmonic analysis; Power measurement; Power system harmonics; Reactive power; Voltage measurement; energy meters verification; measurement; non-sinusoidal conditions; quality assurance;