Preliminary Evaluation of MR Image Quality in a New Clinical ToF-PET/MR System
Author
Delso, Gaspar ; Khalighi, Mehdi ; Hofbauer, Marlena ; Porto, Miguel ; Veit-Haibach, Patrick ; von Schulthess, Gustav
Author_Institution
GE Healthcare´s Global MR Applic. & Workflow Group, Waukesha, WI, USA
Volume
62
Issue
3
fYear
2015
fDate
Jun-15
Firstpage
600
Lastpage
603
Abstract
The goal of the present work was to compare the image quality obtained on a new ToF PET/MR prototype with that of equivalent state-of-the-art standalone systems. The MR image quality of this system was tested by scanning a volunteer with a comprehensive brain protocol. The same exact acquisition was repeated before and after the PET detectors were installed. Furthermore, a baseline measurement was acquired by importing the protocol on a GE Discovery 750 w MR system and repeating the acquisition on the same subject. The obtained datasets were registered and reviewed by medical doctors with experience in both radiology and nuclear imaging. Structure detectability, delineation and noise ratio were considered. MR image quality was shown to be virtually identical between the hybrid (pre- and post-insert) and standalone systems. We conclude that clinical MR sequences are not qualitatively affected by the presence of the PET detector insert.
Keywords
biomedical MRI; brain; image registration; image sequences; medical image processing; positron emission tomography; GE Discovery 750 w MR system; MR image quality; clinical MR sequences; clinical ToF-PET/MR System; comprehensive brain protocol; delineation; image registration; magnetic resonance imaging; noise ratio; nuclear imaging; radiology; structure detectability; Acceleration; Detectors; Image quality; Positron emission tomography; Signal to noise ratio; Tensile stress; Hybrid; PET/MR; time of flight;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2015.2414275
Filename
7110629
Link To Document