• DocumentCode
    722228
  • Title

    Effectiveness of safe working procedure on SMF exposure levels and work performances in 3 T MRI system operations

  • Author

    Yamaguchi-Sekino, S. ; Sekino, M. ; Nakai, T.

  • Author_Institution
    Nat. Inst. of Occupational Safety & Health, Kawasaki, Japan
  • fYear
    2015
  • fDate
    11-15 May 2015
  • Firstpage
    1
  • Lastpage
    1
  • Abstract
    Occupational exposure to the static magnetic field (SMF) is an arising problem in operators of magnetic resonance imaging (MRI) system. It has been well established that human body movements in the stray field from MRI system induces the electric current inside the body and this magnetically induced current causes the temporal sensational changes such as vertigo, headache, nausea and metallic taste etc... to MR workers [1]. Although the self-motivated motion control is recommended for MR workers to prevent these temporal effects [1], more feasible safe working procedure is required to achieve the compliance monitoring. The present study performed the trial of the safe working procedure among 3 T MRI system users and investigated its effectiveness in occupational SMF exposure levels and their work performances.
  • Keywords
    biomechanics; biomedical MRI; diseases; electric current; MRI system; electric current; headache; human body movements; magnetic flux density 3 T; magnetic resonance imaging system; magnetically induced current; metallic taste; nausea; occupational SMF exposure levels; safe working procedure; self-motivated motion control; static magnetic field; temporal sensational changes; vertigo; Geriatrics; Guidelines; Magnetic fields; Magnetic heads; Magnetic resonance imaging; Magnetometers; Motion control;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Magnetics Conference (INTERMAG), 2015 IEEE
  • Conference_Location
    Beijing
  • Print_ISBN
    978-1-4799-7321-7
  • Type

    conf

  • DOI
    10.1109/INTMAG.2015.7157570
  • Filename
    7157570