DocumentCode :
73095
Title :
Rechargeable neurostimulator with H-shaped fractal slots in PCB copper layers
Author :
Qingfeng Li ; Shaobo Chen ; Weiming Wang ; Hongwei Hao ; Luming Li
Author_Institution :
Sch. of Aerosp., Tsinghua Univ., Beijing, China
Volume :
51
Issue :
11
fYear :
2015
fDate :
5 28 2015
Firstpage :
813
Lastpage :
815
Abstract :
Rechargeable neurostimulators make it possible to research and treat brain disorders for a long period of time. However, metal objects such as device cases and inner copper layers of the printed circuit board (PCB) lower the energy transfer efficiency and can even present safety risks due to the eddy currents in these metals. The use of a fractal pattern slot in the metal to suppress the eddy currents is proposed and a rechargeable neurostimulator with deformed H-shape fractal slots in the PCB inner copper layers is described. The results show that the use of an H-shaped fractal pattern slot is effective over a relatively wide frequency range and the maximum temperature of the neurostimulator is in the range of 34.5-39.5°C, which indicates that the device can be used safely for long-term animal studies and even in the clinical treatment of brain disorders.
Keywords :
biomedical electronics; biomedical equipment; brain; copper; eddy currents; fractals; medical disorders; neurophysiology; patient treatment; printed circuits; Cu; H-shaped fractal pattern slot; PCB; PCB copper layers; brain disorders; clinical treatment; eddy currents; energy transfer efficiency; printed circuit board; rechargeable neurostimulator; safety risks; temperature 34.5 degC to 39.5 degC;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el.2014.4424
Filename :
7110773
Link To Document :
بازگشت