RESCUE - Reduction of MRI SNR Degradation by Using an MR-Synchronous Low-Interference PET Acquisition Technique
Author :
Gebhardt, Pierre ; Wehner, Jakob ; Weissler, Bjoern ; Frach, Thomas ; Marsden, Paul K. ; Schulz, Volkmar
Author_Institution :
Div. of Imaging Sci. & Biomed. Eng., King´s Coll. London, London, UK
Volume :
62
Issue :
3
fYear :
2015
fDate :
6/1/2015 12:00:00 AM
Firstpage :
634
Lastpage :
643
Abstract :
Devices aiming at combined Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI) to enable simultaneous PET/MR image acquisition have to fulfill demanding requirements to avoid mutual magneticas well as electromagnetic-field-related interferences which lead to image quality degradation. Particularly Radio-Frequency (RF)-field-related interferences between PET and MRI may lead to MRI SNR reduction, thereby deteriorating MR image quality. RF shielding of PET electronics is therefore commonly applied to reduce RF emission and lower the potential coupling into MRI RF coil(s). However, shields introduce eddy-current-induced MRI field distortions and should thus be minimized or ideally omitted. Although the MRI noise floor increase caused by a PET system might be acceptable for many MRI applications, some MRI protocols, such as fast or high-resolution MRI scans, typically suffer from low SNR and might need more attention regarding RF silence to preserve the intrinsic MRI SNR. For such cases, we propose RESCUE, an MRI-synchronously-gated PET data acquisition technique: By interrupting the PET acquisition during MR signal receive phases, PET-related RF emission may be minimized, leading to MRI SNR preservation. Our PET insert Hyperion IID using Philips Digital Photon Counting (DPC) sensors serves as the platform to demonstrate RESCUE. To make the DPC sensor suitable for RESCUE to be applied for many MRI sequences with acquisition time windows in the range of a few milliseconds, we present in this paper a new technique which enables rapid DPC sensor operation interruption by dramatically lowering the overhead time to interrupt and restart the sensor operation. Procedures to enter and leave gated PET data acquisition may imply sensitivity losses which add to the ones occurring during MRI RF acquisition. For the case of our PET insert, the new DPC quick-interruption technique yields a PET sensitivity loss reduction by a factor of 78 when compared - o the loss introduced with the standard start/stop procedure. For instance, PET sensitivity losses related to overhead time are 2.9% in addition to the loss related to PET gating being equal to the MRI RF acquisition duty cycle (14.7%) for an exemplary T1-weighted 3D-FFE MRI sequence. MRI SNR measurement results obtained with one Singles Detection Module (SDM) using no RF shield demonstrate a noise floor reduction by a factor of 2.1, getting close to the noise floor level of the SNR reference scan (SDM off-powered) when RESCUE was active.
Keywords :
biomedical MRI; biomedical equipment; data acquisition; electromagnetic interference; photodetectors; positron emission tomography; DPC quick interruption technique; Hyperion IID; MR signal receive phases; MR synchronous PET acquisition technique; MRI RF acquisition duty cycle; MRI RF coils; MRI SNR degradation reduction; MRI SNR measurement; MRI noise floor; MRI synchronously gated PET data acquisition technique; PET acquisition interruption; PET electronics; PET related RF emission; PET sensitivity loss reduction; Philips DPC sensors; Philips Digital Photon Counting sensors; RESCUE; RF shielding; T1 weighted 3D-FFE MRI sequence; eddy current induced MRI field distortions; electromagnetic field related interferences; fast MRI scans; high resolution MRI scans; low interference PET acquisition technique; magnetic resonance imaging; positron emission tomography; radiofrequency field related interferences; rapid DPC sensor operation interruption; sensor acquisition time windows; simultaneous PET-MR image acquisition; singles detection module; Data acquisition; Magnetic resonance imaging; Positron emission tomography; Radio frequency; Sensors; Signal to noise ratio; Digital; FPGA; RF interference reduction; RF silence; SNR; dSiPM; shielding; simultaneous PET/MRI;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2015.2434851
Filename :
7117461
Link To Document :
بازگشت