• DocumentCode
    747969
  • Title

    New calibration method for coaxial microcalorimeters

  • Author

    Brunetti, Luciano ; Vremera, Emil T.

  • Author_Institution
    Inst. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
  • Volume
    54
  • Issue
    2
  • fYear
    2005
  • fDate
    4/1/2005 12:00:00 AM
  • Firstpage
    684
  • Lastpage
    687
  • Abstract
    The most critical point in the realization of an electrical power standard based on a microcalorimeter is the accurate determination of the microcalorimeter losses and their effect on the measurand, i.e., the effective efficiency of the sensor under calibration. At the Istituto Elettrotecnico Nazionale (IEN) Galileo Ferraris, a procedure has been introduced that makes the microcalorimeter constant determination more accurate and reliable. Experiments with the 7-mm coaxial microcalorimeter in 50 MHz-18 GHz range have shown a significant improvement in the uncertainty for the Italian national power standard at the 0.2% level.
  • Keywords
    UHF measurement; calibration; calorimeters; coaxial cables; measurement standards; microwave measurement; power measurement; 0.50 to 18 GHz; 7 mm; Istituto Elettrotecnico Nazionale Galileo Ferraris; Italian national power standard; calibration method; coaxial microcalorimeters; electrical power standard; microcalorimeter constant determination; microcalorimeter losses; sensor efficiency; Calibration; Coaxial components; Electric variables measurement; Hafnium; Insulation; Loss measurement; Power measurement; Thermal loading; Thermal sensors; Thermistors; Microcalorimeter; microwave; power; standard;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2005.843555
  • Filename
    1408264