DocumentCode
748259
Title
Quartz planar multijunction thermal converter as a new AC-DC current transfer standard up to 1 MHz
Author
Scarioni, Luciana ; Klonz, Manfred ; Funck, Torsten
Author_Institution
Phys. Tech. Bundesanstalt, Braunschweig, Germany
Volume
54
Issue
2
fYear
2005
fDate
4/1/2005 12:00:00 AM
Firstpage
799
Lastpage
802
Abstract
The new generation of PMJTCs on quartz crystal chips and on quartz membranes developed at the Physikalisch-Technische Bundesanstalt (PTB) results in new calculable ac-dc current transfer standards for a current of 10 mA in the frequency range from 10 kHz to 1 MHz, with considerably reduced uncertainties. AC-DC current transfer differences below 1 μA/A with standard uncertainties below 1 μA/A have been achieved.
Keywords
convertors; electric current measurement; frequency measurement; measurement uncertainty; quartz; transfer standards; 0.010 to 1 MHz; 10 mA; AC-DC current transfer standard; Physikalisch-Technische Bundesanstalt; measurement uncertainties; planar multijunction thermal converter; quartz crystal chips; quartz membranes; standard uncertainties; thermal converters; Biomembranes; Coaxial components; Frequency; Heat transfer; Helium; Impedance; Resistance heating; Standards development; Thermal resistance; Uncertainty; AC–DC current transfer; quartz-crystal; thermal converters;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2004.843352
Filename
1408292
Link To Document