Estimating linear functionals of a PET image
Author
Bickel, P.J. ; Ritov, Y.
Author_Institution
Dept. of Stat., California Univ., Berkeley, CA, USA
Volume
14
Issue
1
fYear
1995
fDate
3/1/1995 12:00:00 AM
Firstpage
81
Lastpage
87
Abstract
The authors discuss the effect of having a finite number of detectors on the estimation of a bounded linear functional of a PET image. They argue that the functional should be estimated directly from the detector counts and give an estimate of the effect of the binning of the counts to order D-2, where D is the number of detectors. The authors then suggest an estimator that includes a data determined bias correction reducing the bias to o(D-2)
Keywords
functional equations; positron emission tomography; PET image; bounded linear functional; counts binning; data determined bias correction; detector counts; diagnostic nuclear medicine; finite detectors number; linear functionals estimation; medical diagnostic imaging; Biochemistry; Biomedical imaging; Gamma ray detection; Gamma ray detectors; Gamma rays; Image reconstruction; Positron emission tomography; Radioactive materials; Statistics; Sugar;