    Microstrip-antenna design for hyperthermia treatment of superficial tumors

  • Author

    Montecchia, Francesco

  • Author_Institution
    Dept. of Electron. & Autom., Ancona Univ., Italy
  • Volume
    39
  • Issue
    6
  • fYear
    1992
  • fDate
    6/1/1992 12:00:00 AM
  • Firstpage
    580
  • Lastpage
    588
  • Abstract
    Three microstrip antennas of increasing complexity with electromagnetic and heating characteristics potentially suitable as applicators for superficial hyperthermia have been designed, developed, and tested in different radiative conditions: a microstrip disk, a microstrip annular-slot, and a microstrip spiral. Electromagnetic design criteria are presented together with the determinations of the applicator return loss versus frequency and thermograms of the near-field heating pattern in muscle-like phantom. The results are in good agreement with theory.
  • Keywords
    biomedical equipment; biothermics; microstrip antennas; radiation therapy; applicator return loss; electromagnetic design criteria; heating characteristics; hyperthermia treatment; microstrip annular-slot; microstrip disk; microstrip spiral; muscle-like phantom; near-field heating pattern; radiative conditions; superficial hyperthermia; superficial tumors; thermograms; Applicators; Electromagnetic heating; Electromagnetic radiation; Frequency; Hyperthermia; Imaging phantoms; Microstrip antennas; Neoplasms; Spirals; Testing; Electromagnetics; Equipment Design; Hyperthermia, Induced; Neoplasms;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.141196
  • Filename
    141196