DocumentCode :
753617
Title :
Closed-loop class E transcutaneous power and data link for MicroImplants
Author :
Troyk, Philip R. ; Schwan, Martin A K
Author_Institution :
Pritzker Inst. of Med. Eng., Illinois Inst. of Technol., Chicago, IL, USA
Volume :
39
Issue :
6
fYear :
1992
fDate :
6/1/1992 12:00:00 AM
Firstpage :
589
Lastpage :
599
Abstract :
The use of a multifrequency transmitter coil driver based on the class E topology is described. The development of a high-Q approximation, which simplifies the design procedure is presented. A closed-loop controller to compensate for transmitter and receiver variations, and a method of data modulation using synchronous frequency shifting are described. The closed-loop class E circuit shows great promise, especially for circuits with unusually low coefficients of coupling. Currents of several amperes, at radio frequencies, can easily and efficiently be obtained.
Keywords :
prosthetic power supplies; prosthetics; MicroImplants; class E topology; closed-loop controller; data modulation method; high-Q approximation; multifrequency transmitter coil driver; synchronous frequency shifting; transcutaneous power/data link; Circuit topology; Coils; Coupling circuits; Driver circuits; Implants; Needles; Power supplies; Radio frequency; Radio transmitters; Telemetry; Equipment Design; Prostheses and Implants; Telemetry; Transcutaneous Electric Nerve Stimulation;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.141197
Filename :
141197
Link To Document :
بازگشت