• DocumentCode
    753617
  • Title

    Closed-loop class E transcutaneous power and data link for MicroImplants

  • Author

    Troyk, Philip R. ; Schwan, Martin A K

  • Author_Institution
    Pritzker Inst. of Med. Eng., Illinois Inst. of Technol., Chicago, IL, USA
  • Volume
    39
  • Issue
    6
  • fYear
    1992
  • fDate
    6/1/1992 12:00:00 AM
  • Firstpage
    589
  • Lastpage
    599
  • Abstract
    The use of a multifrequency transmitter coil driver based on the class E topology is described. The development of a high-Q approximation, which simplifies the design procedure is presented. A closed-loop controller to compensate for transmitter and receiver variations, and a method of data modulation using synchronous frequency shifting are described. The closed-loop class E circuit shows great promise, especially for circuits with unusually low coefficients of coupling. Currents of several amperes, at radio frequencies, can easily and efficiently be obtained.
  • Keywords
    prosthetic power supplies; prosthetics; MicroImplants; class E topology; closed-loop controller; data modulation method; high-Q approximation; multifrequency transmitter coil driver; synchronous frequency shifting; transcutaneous power/data link; Circuit topology; Coils; Coupling circuits; Driver circuits; Implants; Needles; Power supplies; Radio frequency; Radio transmitters; Telemetry; Equipment Design; Prostheses and Implants; Telemetry; Transcutaneous Electric Nerve Stimulation;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.141197
  • Filename
    141197