DocumentCode
75482
Title
Performance Characteristics of the OSMOND Neutron Detector
Author
Bateman, J.E. ; Charlton, T.R. ; Dalgliesh, R. ; Duxbury, D.M. ; Kinane, C.J. ; Rhodes, N.J. ; Schooneveld, E.M. ; Spill, E.J.
Author_Institution
Technol. Dept., Rutherford Appleton Lab., Oxford, UK
Volume
60
Issue
3
fYear
2013
fDate
Jun-13
Firstpage
2232
Lastpage
2236
Abstract
The performance characteristics of the Off Specular MicrOstrip Neutron Detector (OSMOND) are described. Based on microstrip gas chamber (MSGC) detectors, OSMOND is a high counting rate detector, with sub-millimeter position resolution, capable of replacing the existing rate limited scintillator detectors currently in use on the suite of reflectometer instruments of the ISIS spallation neutron source. The detector system is described together with results of neutron beam tests carried out on the CRISP and POLREF reflectometers.
Keywords
ionisation chambers; neutron detection; neutron sources; particle beam diagnostics; proportional counters; scintillation counters; CRISP reflectometer; ISIS spallation neutron source; OSMOND neutron detector; Off Specular MicrOstrip Neutron Detector; POLREF reflectometer; high counting rate detector; limited scintillator detectors; microstrip gas chamber detectors; neutron beam tests; reflectometer instrument suite; sub-millimeter position resolution; Detectors; Instruments; Microstrip; Neutrons; Nickel; Particle beams; Spatial resolution; $^{3}$ He neutron detector; Micropattern gas detector;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2013.2260560
Filename
6519336
Link To Document