Abstract :
A counting-mode line detector has been evaluated at a synchrotron radiation source in order to assess its performance for imaging applications. The x-ray detector is based on 3 mm thick CdZnTe arrays with 1 mm pixel pitch and multi-threshold counting electronics. Data readout has been performed in threshold scan mode to provide maximum energy dispersion. The acquired energy spectra are interpreted regarding the characteristics of a periodic source and pulse pile-up in the detector. In particular, energy response and rate behavior of single-photon and double-photon events are investigated. Count rate dynamics are studied up to incident rates of . A further subject of the investigation is the variance of the observed counting rate. Linearity in the energy domain is tested between 40 keV and 140 keV.
Keywords :
X-ray apparatus; position sensitive particle detectors; readout electronics; semiconductor counters; synchrotron radiation; CdZnTe arrays; X-ray detector; data readout; double-photon events; energy spectra; energy-dispersive counting-mode line detector; single-photon events; synchrotron radiation source; Biomedical imaging; Europe; Optical imaging; Radiation detectors; Storage rings; Synchrotron radiation; Testing; X-ray detection; X-ray detectors; X-ray imaging; Coincidence; photon counting; pulse pile-up; synchrotron; x-ray detector; x-ray imaging;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2008.924083
Filename :
4545071
Link To Document :
بازگشت