• DocumentCode
    768094
  • Title

    Open Type of Magnetically Shielded Room Combined With Cancelling Coil for Magnetic Resonance Imaging

  • Author

    Hirosato, Shigetaka ; Yamazaki, Keita ; Muramatsu, Kazuhiro ; Hirayama, Masayuki ; Haga, Akira ; Katada, Kazuhiro

  • Author_Institution
    R & D Inst., Takenaka Corp
  • Volume
    42
  • Issue
    10
  • fYear
    2006
  • Firstpage
    3542
  • Lastpage
    3544
  • Abstract
    In order to improve amenity for patients in hospitals, we have designed an open type of magnetically shielded room (MSR) for magnetic resonance imaging combined with cancelling coils, which enable transparent opening. First, an optimal arrangement of cancelling coils is investigated using three-dimensional magnetic field analysis. Secondly, the validity of the analysis method is verified and the realization of an open type of MSR designed both for keeping low leakage flux density and for maximizing the opening is confirmed by experiments using a small MSR model
  • Keywords
    biomedical MRI; magnetic fields; magnetic shielding; patient diagnosis; 3D magnetic field analysis; cancelling coil; flux density; magnetic resonance imaging; magnetically shielded room; patient diagnosis; transparent opening; Bars; Coils; Computed tomography; Hospitals; Magnetic analysis; Magnetic field measurement; Magnetic flux; Magnetic flux leakage; Magnetic resonance imaging; Magnetic shielding; Amenity of patients; cancelling coil; magnetic field analysis; magnetic resonance imaging; magnetically shielded room;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.2006.883835
  • Filename
    1704687