• DocumentCode
    768203
  • Title

    Improvement of the Transcutaneous Energy Transmission System Utilizing Ferrite Cored Coils for Artificial Hearts

  • Author

    Miura, Hidekazu ; Arai, Shinsuke ; Kakubari, Yasuyuki ; Sato, Fumihiro ; Matsuki, Hidetoshi ; Sato, Tadakuni

  • Author_Institution
    Graduate Sch. of Eng., Tohoku Univ.
  • Volume
    42
  • Issue
    10
  • fYear
    2006
  • Firstpage
    3578
  • Lastpage
    3580
  • Abstract
    Newly designed transcutaneous energy transmission system for a ventricular assist device was developed. We obtained higher dc-dc efficiency of 93.4% and smaller size of devices. Also, the devices were designed in view of the biomedical compatibility, gradually sloped coils and the flexibility of the primary coil prevent pressure necroses of the skin. The rectifier circuit board was placed in the internal space of the ferrite core to reduce the number of the implanted devices. However, radiated outward ground patterns were adopted to achieve a balance between thermal radiation and reduction of eddy-current loss
  • Keywords
    artificial organs; coils; ferrites; heat radiation; prosthetic power supplies; artificial hearts; contactless power supply; ferrite cored coils; pressure necroses; rectifier circuit board; synchronous rectification; thermal radiation; transcutaneous energy transmission; Artificial heart; Coils; Ferrites; Flexible printed circuits; Magnetic cores; Magnetic resonance; Phase locked loops; Printed circuits; Rectifiers; Voltage; Artificial heart; contactless power supply; digital phase-locked loop (PLL); synchronous rectification;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.2006.879630
  • Filename
    1704698