  • Title

    Echo Capsule for Medical Use (A Batteryless Endoradiosonde)

  • Author

    Nagumo, J. ; Uchiyama, A. ; Kimoto, S. ; Watanuki, T. ; Hori, M. ; Suma, K. ; Ouchi, A. ; Kumano, M. ; Watanabe, H.

  • Author_Institution
    Department of Applied Physics, Faculty of Engineering, University of Tokyo, Japan
  • Volume
    9
  • Issue
    3
  • fYear
    1962
  • fDate
    7/1/1962 12:00:00 AM
  • Firstpage
    195
  • Lastpage
    199
  • Abstract
    A new type of a swallowable radio transmitter, ECHO CAPSULE, is described along with some clinical experiments. Unlike other similar capsules which have been reported, this capsule contains no battery and is energized from outside the human body. The energy is converted into dc by the transistor of the oscillator circuit itself, stored into a capacitor and then used as an energy source for the transmission of information on temperature, acidity or the like within the digestive tract.