DocumentCode :
769839
Title :
Restoration of gamma camera-based nuclear medicine images
Author :
Hon, T.C. ; Rangayyan, Rangaraj M. ; Hahn, L.J. ; Kloiber, R.
Author_Institution :
Dept. of Electr. Eng., Calgary Univ., Alta., Canada
Volume :
8
Issue :
4
fYear :
1989
fDate :
12/1/1989 12:00:00 AM
Firstpage :
354
Lastpage :
363
Abstract :
Experiments were conducted using a Siemens Rota camera to study the applicability of two linear shift-invariant (LSI) filters, namely, the Wiener and power spectrum equalization filters, for restoration of planar projections and single-photon-emission computed tomography (SPECT) images. In the restoration scheme, the system transfer function, computed from a line source image, is modeled by a 2-D Gaussian function. The noise power spectrum is modeled as a constant for planar images and as a ramp for SPECT images. The filters have been applied to restore computer-simulated 1-D and 2-D projections and SPECT images of two simple phantoms, 2-D projections of two phantoms obtained from the Siemens Rota camera, and SPECT images of a cardiac phantom obtained from the Siemens Rota camera. The filters are shown to perform partial restoration. Considerable noise suppression and detail enhancement have been observed in the restored images. quantitative measurements such as root-mean-squared error and contrast ratio have been used for objective analysis of the results, which are encouraging
Keywords :
picture processing; radioisotope scanning and imaging; 2D Gaussian function; Siemens Rota camera; Wiener filter; cardiac phantom; contrast ratio; detail enhancement; gamma camera-based nuclear medicine images; image restoration; linear shift-invariant filter; medical diagnostic imaging; noise power spectrum; noise suppression; planar projections; power spectrum equalisation filter; root-mean-squared error; single-photon-emission computed tomography; Cameras; Image restoration; Imaging phantoms; Large scale integration; Nonlinear filters; Nuclear medicine; Power filters; Power system modeling; Power system restoration; Wiener filter;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.41488
Filename :
41488
Link To Document :