Title :
A Low Level Beta Scintillation Counter
Author :
Cassidy, Melvin E. ; Graveson, Robert T.
Author_Institution :
Health and Safety Laboratory U. S. Atomic Energy Commission New York, New York
fDate :
7/1/1963 12:00:00 AM
Abstract :
An automatic scintillation counter for the measurement of low beta activity samples is described. Background is in the order of one (1) count per minute. Efficiency of counting for K-40 betas is thirty (30) per cent. The scintillator is a thin plastic phosphor mounted on the sample which makes a simple, rugged package suitable for automatic sample changing. Photomultiplier tube noise measurements, preamplifier circuit, and the mechanical details are discussed.
Keywords :
Atomic measurements; Contamination; Health and safety; Laboratories; Optical coupling; Phosphors; Photomultipliers; Plastics; Radiation detectors; Scintillation counters;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1963.4323287