DocumentCode :
773459
Title :
Evaluation of New Photomultipliers for Scintillation Counting
Author :
Krall, H.R.
Author_Institution :
Radio Corporation of America Electronic Components and Devices Lancaster, Pa.
Volume :
12
Issue :
1
fYear :
1965
Firstpage :
39
Lastpage :
45
Abstract :
This paper presents a comprehensive evaluation of a new bi-alkaliphotocathode photomultiplier developed specifically for scintillation counting. This recently developed photomultiplier is compared to existing photomultiplier types, and data are presented showing photocathode quantum efficiency as a function of wavelength, pulse-height resolution for Cs137 and Fe55, dark-noise spectra at 25 C and -100 C, anode rise time and pulse shape, and pulse-height stability with respect to time and count rate. A method is described for measuring the spatial uniformity of photomultipliers, utilizing a dot-pattern-scanned cathode-ray-tube light source, a pulse-height analyzer, and a live cathode-ray-tube display.
Keywords :
Anodes; Cathodes; Light sources; Photomultipliers; Pulse measurements; Pulse shaping methods; Shape; Spatial resolution; Stability; Wavelength measurement;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1965.4323495
Filename :
4323495