• DocumentCode
    781247
  • Title

    Cutting wax with an ultrasonically activated blade

  • Author

    Vaitekunas, Jeffrey J. ; Stulen, Foster B. ; Grood, Edward S.

  • Author_Institution
    Electr. Eng., Purdue Univ., Hammond, IN
  • Volume
    27
  • Issue
    4
  • fYear
    2008
  • Firstpage
    71
  • Lastpage
    79
  • Abstract
    In this paper, a model was developed for cutting through wax using an ultrasonic instrument in the shear mode. The model was based on viscous heating in a liquid layer of molten wax and extrusion of the molten wax away from an advancing blade, using an energy-balance approach. The model provides a set of equations representing cut rate as a function of velocity and normal force. Using two parameters in the model determined by a regression of experimental data at a single frequency, the model accurately reflected the behavior of experimental data at other frequencies and using blades of different geometries.
  • Keywords
    biomedical ultrasonics; extrusion; waxes; energy-balance approach; extrusion; instrument performance; shear mode; ultrasonically activated blade; viscous heating; wax cutting; Acoustic devices; Blades; End effectors; Heating; Piezoelectric transducers; Predictive models; Shape; Surgery; Surgical instruments; Ultrasonic transducers;
  • fLanguage
    English
  • Journal_Title
    Engineering in Medicine and Biology Magazine, IEEE
  • Publisher
    ieee
  • ISSN
    0739-5175
  • Type

    jour

  • DOI
    10.1109/MEMB.2008.924848
  • Filename
    4558141