DocumentCode :
78132
Title :
Deep brain stimulation lead design to reduce radio-frequency heating in MRI
Author :
Changqing Jiang ; Xiaolong Mo ; Jianqi Ding ; Yantao Dong ; Feng Zhang ; Hongwei Hao ; Luming Li
Author_Institution :
Nat. Eng. Lab. for Neuromodulation, Tsinghua Univ., Beijing, China
Volume :
50
Issue :
25
fYear :
2014
fDate :
12 4 2014
Firstpage :
1898
Lastpage :
1900
Abstract :
Radio-frequency (RF) induced heating at the electrodes of the deep brain stimulation lead is a major concern when used in magnetic resonance imaging (MRI). A novel lead design is proposed to address this problem. By adopting a coil with non-uniform diameters, the maximal temperature rise measured in a standard phantom is reduced from 4.6 to 2.0°C, with an average reduction of 59%. The suppression of RF heating is probably achieved because of the alteration of the transmission line parameters of the lead by the novel design. The details of the design, experiment validation and discussion of the mechanisms are presented.
Keywords :
biomedical MRI; brain; phantoms; MRI; RF heating; deep brain stimulation; electrodes; magnetic resonance imaging; nonuniform diameters; phantom; radiofrequency-induced heating; transmission line parameters;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el.2014.3482
Filename :
6975687
Link To Document :
بازگشت