• DocumentCode
    782412
  • Title

    Vessel wall detection and blood noise reduction in intravascular ultrasound imaging

  • Author

    Gronningsaeter, Aage ; Angelsen, Bjorn A J ; Heimdal, Andreas ; Torp, Hans G.

  • Author_Institution
    Dept. of Biomed. Eng., Trondheim Univ., Norway
  • Volume
    43
  • Issue
    3
  • fYear
    1996
  • fDate
    5/1/1996 12:00:00 AM
  • Firstpage
    359
  • Lastpage
    369
  • Abstract
    Scattering from blood limits the contrast between the vessel wall and the lumen in intravascular ultrasound imaging. This makes it difficult to localize the vessel wall, especially on still images. This paper presents a method for automatic detection of vessel walls and reduction of blood noise based on correlation of the RF-signal between adjacent frames. The ultrasound RF-signal is quadrature demodulated, digitized, stored in memory, and transferred to a computer for processing and analysis. The absolute value of the cross-correlation coefficient between two adjacent frames is used to differentiate between stationary and fluctuating signals. Models and numerical calculations presented in this work indicate that the cross-correlation coefficient obtained from a radially dilating vessel wall will be larger than 0.8 under standard 20 MHz imaging conditions. The corresponding value from blood is less than 0.2 for blood velocities exceeding 0.5 cm s/sup -1/. The blood-noise filter is based on detecting this difference in correlation and displays vessel wall regions with no modifications, while regions detected as blood are rejected. A simplified vessel-wall detector that is suitable for real-time implementation is proposed. The performance of this detector and the blood noise filter are demonstrated by in vitro experiments.
  • Keywords
    biomedical ultrasonics; medical image processing; noise; 20 MHz; RF-signal correlation; blood noise reduction; blood-noise filter; cross-correlation coefficient; intravascular ultrasound imaging; lumen; medical diagnostic imaging; numerical calculations; radially dilating vessel wall; still images; vessel wall detection; Blood; Cells (biology); Detectors; Displays; Filters; Frequency; Noise level; Noise reduction; Spatial resolution; Ultrasonic imaging;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.489392
  • Filename
    489392