DocumentCode
782859
Title
VXI electronics for EUROGAM clover detectors
Author
Lazarus, I. ; Coleman-Smith, P.J. ; Thornhill, J. ; Bosson, G. ; Karkour, N. ; Richard, A. ; Zojceski, Z. ; Ring, C.
Author_Institution
SERC Daresbury Lab., Warrington, UK
Volume
42
Issue
6
fYear
1995
fDate
12/1/1995 12:00:00 AM
Firstpage
2288
Lastpage
2291
Abstract
A single-width D-sized VXI card has been designed and built to fully instrument one of the new Clover detectors designed for experiments with the EUROGAM array. Each Clover detector consists of four individual germanium (Ge) crystals and an escape suppression shield comprising 16 bismuth germanate (BGO) crystals. The Clover electronics card measures energy and timing data to 13-bit accuracy from each Ge crystal for spectroscopy and measures sum energy and timing for the BGO shield. It also records a complete Ge and BGO hit pattern for each event and performs intelligent hardware escape suppression to exploit the segmented nature of the Clover detectors
Keywords
detector circuits; germanium radiation detectors; nuclear electronics; solid scintillation detectors; BGO crystal; Bi4Ge3O12; Clover detectors; EUROGAM; Ge; Ge crystal; VXI card; VXI electronics; escape suppression shield; intelligent hardware escape suppression; Bismuth; Crystals; Detectors; Energy measurement; Germanium; Hardware; Instruments; Sensor arrays; Spectroscopy; Timing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.489429
Filename
489429
Link To Document