DocumentCode :
783395
Title :
Automatic compensation system for impedance measurement
Author :
Carullo, Alessio ; Parvis, Marco ; Vallan, Alberto ; Callegaro, Luca
Author_Institution :
Dipt. di Elettronica, Politecnico di Torino, Italy
Volume :
52
Issue :
4
fYear :
2003
Firstpage :
1239
Lastpage :
1242
Abstract :
This paper deals with the realization of the four-pair terminal definition of impedance standards. A simple, though reliable, system is described that allows an automatic compensation of the voltage at the low potential port of impedance standards to be obtained. Such a system employs a commercial data acquisition board and a signal generator with adjustable-phase capability, which acts as the phase reference for the generator that feeds the impedance standard. A standard PC controls the whole system and implements the demodulation and the control algorithms. Preliminary tests have been performed in the frequency range of 50 Hz to 20 kHz with different kinds of impedance standards (resistive, inductive and capacitive), obtaining a residual voltage at the low potential port of less than 5 μV.
Keywords :
compensation; electric impedance measurement; measurement standards; multiterminal networks; 50 Hz to 20 kHz; automatic compensation system; control algorithm; data acquisition; demodulation algorithm; four-terminal pair network; impedance measurement; impedance standard; personal computer; phase reference; signal generator; Automatic control; Control systems; Data acquisition; Demodulation; Feeds; Impedance measurement; Low voltage; Performance evaluation; Signal generators; Testing;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2003.816843
Filename :
1232374
Link To Document :
بازگشت