DocumentCode :
784151
Title :
Micromachined Hot-Wire Thermal Conductivity Probe for Biomedical Applications
Author :
Ming Yi ; Panchawagh, Hrishikesh V. ; Podhajsky, Ronald J. ; Mahajan, Roop L.
Author_Institution :
Critical Technol. & Appl. Sci., Virginia Tech, Blacksburg, VA, USA
Volume :
56
Issue :
10
fYear :
2009
Firstpage :
2477
Lastpage :
2484
Abstract :
This paper presents the design, fabrication, numerical simulation, and experimental validation of a micromachined probe that measures thermal conductivity of biological tissues. The probe consists of a pair of resistive line heating elements and resistance temperature detector sensors, which were fabricated by using planar photolithography on a glass substrate. The numerical analysis revealed that the thermal conductivity and diffusivity can be determined by the temperature response induced by the uniform heat flux in the heating elements. After calibrating the probe using a material (agar gel) of known thermal conductivity, the probe was deployed to calculate the thermal conductivity of Crisco. The measured value is in agreement with that determined by the macro-hot-wire probe method to within 3%. Finally, the micro thermal probe was used to investigate the change of thermal conductivity of pig liver before and after RF ablation treatment. The results show an increase in thermal conductivity of liver after the RF ablation.
Keywords :
bioMEMS; biological tissues; biothermics; calibration; gels; liver; numerical analysis; photolithography; temperature sensors; thermal conductivity; thermal diffusivity; RF ablation treatment; agar gel; biological tissues; biomedical applications; calibration; diffusivity; glass substrate; heat flux; macro-hot-wire probe method; micromachined probe; microthermal probe; numerical simulation; pig liver; planar photolithography; resistance temperature detector sensors; resistive line heating elements; thermal conductivity; Biomedical measurements; Conductivity measurement; Electrical resistance measurement; Fabrication; Liver; Numerical simulation; Probes; Radio frequency; Temperature sensors; Thermal conductivity; Microprobe; RF ablation; sensor; thermal conductivity; tissue; Algorithms; Animals; Calibration; Catheter Ablation; Computer Simulation; Equipment Design; Liver; Microtechnology; Models, Theoretical; Reproducibility of Results; Swine; Temperature; Thermal Conductivity; Thermography; Thermometers;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2009.2020991
Filename :
4895286
Link To Document :
بازگشت