Title :
A method for improved standardization of in vivo calcaneal time-domain speed-of-sound measurements
Author_Institution :
Center for Devices & Radiol. Health, Silver Spring, MD
fDate :
7/1/2008 12:00:00 AM
Abstract :
Although calcaneal speed of sound (SOS) is an effective predictor of osteoporotic fracture risk, clinical SOS measurements exhibit a high degree of inter-system variability. Calcaneal SOS is usually computed from time-of-flight measurements of broadband ultrasound pulses that propagate through the foot. In order to minimize the effects of multi-path interference, many investigators measure time-of-flight from markers near the leading edge of the pulse. The calcaneus is a highly attenuating, highly inhomogeneous bone that distorts propagating ultrasound pulses via frequency-dependent attenuation, reverberation, dispersion, multiple scattering, and refraction. This pulse distortion can produce errors in leading-edge transit-time marker-based SOS measurements. In this paper, an equation to predict dependence of time-domain SOS measurements on system parameters (center frequency and bandwidth), transit-time marker location, and bone properties (attenuation coefficient and thickness) is validated with through-transmission measurements in a bone-mimicking phantom and in 73 women in vivo, using a clinical bone sonometer. In order to test the utility of the formula for suppressing system dependence of SOS measurements, a wideband laboratory data acquisition system was used to make a second set of through-transmission measurements on the phantom. The compensation formula reduced system-dependent leading-edge transit-time marker-based SOS measurements in the phantom from 41 m/s to 5 m/s and reduced average transit-time marker-related SOS variability in 73 women from 40 m/s to 10 m/s. The compensation formula can be used to improve standardization in bone sonometry.
Keywords :
bioacoustics; biomechanics; biomedical measurement; biomedical ultrasonics; bone; diseases; fracture; inhomogeneous media; phantoms; time-domain analysis; ultrasonic absorption; ultrasonic dispersion; ultrasonic propagation; ultrasonic transmission; bone properties; bone-mimicking phantom; broadband ultrasound pulse; calcaneal time-domain speed-of-sound measurement; clinical SOS measurement; clinical bone sonometer; data acquisition system; dispersion; frequency-dependent attenuation; inhomogeneous bone; inter-system variability; leading-edge transit-time marker; multipath interference; multiple scattering; osteoporotic fracture risk prediction; pulse distortion; refraction; reverberation; time-of-flight from markers; time-of-flight measurement; transmission measurement; Bones; Distortion measurement; Imaging phantoms; In vivo; Pulse measurements; Standardization; Time domain analysis; Ultrasonic imaging; Ultrasonic variables measurement; Velocity measurement; Calcaneus; Elastic Modulus; Elasticity Imaging Techniques; Female; Humans; Image Interpretation, Computer-Assisted; Middle Aged; Reproducibility of Results; Sensitivity and Specificity; Stress, Mechanical; Ultrasonography;
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
DOI :
10.1109/TUFFC.2008.822