Abstract
The present paper describes an improved version of a four-standard calibration procedure for six-port reflectometers that utilizes simple iteration computations instead of the other more cumbersome optimization techniques attempted previously. The four standards required are a matched load (with a return loss not more than -40 dB) and three highly reflecting terminations (each with a return loss not less than -2 dB)
Keywords
calibration; iterative methods; measurement standards; microwave reflectometry; reflectometers; -2 dB; -40 dB; calibration; four-standard procedure; iteration; matched load; optimization; reflecting terminations; return loss; six-port reflectometers; Calibration; Detectors; Hardware; Instruments; Measurement standards; Nonlinear equations; Particle measurements; Reflection; Software standards; Testing;