Title :
Ultrasonic texture motion analysis: theory and simulation
Author :
Meunier, Jean ; Bertrand, Michel
Author_Institution :
Dept. d´´Inf. et de Recherche Oper., Montreal Univ., Que., Canada
Volume :
14
Issue :
2
fYear :
1995
fDate :
6/1/1995 12:00:00 AM
Firstpage :
293
Lastpage :
300
Abstract :
A theoretical model was previously developed to evaluate the relationship between the dynamics of ultrasonic speckle and its underlying tissue. The model is divided into an instrumental part represented by the point spread function (in the far field) of the ultrasonic apparatus and a moving tissue component described by a collection of scatterers. By computing the convolution of these terms and then the envelope, one obtains a simulated ultrasonic speckle pattern sequence which shows speckle motions closely linked to the tissue dynamics when small motion amplitudes are involved. Here, a theoretical study of the correlation between various linear transformations of the tissue and the corresponding ultrasonic speckle motions is performed, based on a 2D extension of the envelope cross-correlation analysis of a narrow-band Gaussian noise. In the linear scan case, obviously, tissue translation generates an identical speckle translation. However, tissue/speckle motion correlation decreases with increasing rotation and/or biaxial deformation, lateral deformation (perpendicular to the beam propagation axis) being much less sensitive. With respect to the transducer frequency, the rotation and the axial deformation of the tissue show a better relationship with their respective speckle motion at lower frequencies while lateral deformation correlation is independent of the pulse frequency. With respect to beam (pulse) size parameters, tissue/speckle correlation decreases with rotation when a wide ultrasonic beam is used while the axial deformation correlation decreases with the axial duration of the pulse. This study sets the ground for the development of an ultrasonic strain gauge particularly useful for the assessment of biomechanical soft tissue and fluid flow properties based on speckle tracking
Keywords :
biomedical ultrasonics; image texture; medical image processing; motion estimation; speckle; biaxial deformation; biomechanical soft tissue properties; envelope cross-correlation analysis; fluid flow properties; lateral deformation; moving tissue component; narrow-band Gaussian noise; point spread function; speckle tracking; terms convolution; tissue dynamics; tissue linear transformations; tissue translation; ultrasonic strain gauge; ultrasonic texture motion analysis; Analytical models; Computational modeling; Convolution; Frequency; Instruments; Motion analysis; Narrowband; Performance analysis; Scattering; Speckle;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.387711
Filename :
387711
Link To Document :
بازگشت