Abstract
The authors describe the implementation of a maximum likelihood (ML) algorithm using expectation maximization (EM) for pinhole SPECT with a displaced center-of-rotation. A ray-tracing technique is used in implementing the ML-EM algorithm. The proposed ML-EM algorithm is able to correct the center of rotation displacement which can be characterized by two orthogonal components. The algorithm is tested using experimentally acquired data, and the results demonstrate that the pinhole ML-EM algorithm is able to correct artifacts associated with the center-of-rotation displacement
Keywords
image reconstruction; medical image processing; single photon emission computed tomography; artifacts correction; center-of-rotation displacement; displaced center-of-rotation; experimentally acquired data; maximum likelihood reconstruction; medical diagnostic imaging; nuclear medicine; orthogonal components; pinhole SPECT; ray-tracing technique; Geometry; Image reconstruction; Image resolution; Iterative methods; Optical collimators; Ray tracing; Reconstruction algorithms; Single photon emission computed tomography; Testing; Time measurement;