    Paralleling Planar Ge(Li) Detectors for Counting Large Volume Biological Samples

  • Author

    Saunders, E.W. ; Maxwell, C.J.

  • Author_Institution
    Electronics Engineering Research Division
  • Volume
    15
  • Issue
    1
  • fYear
    1968
  • Firstpage
    423
  • Lastpage
    431
  • Abstract
    Four planar Ge(Li) detectors have been operated in an array, in parallel electrically, and connected to a single preamplifier. The total active volume is 28 cm3. The photopeak efficiency of the four detector array at 1.33 MeV is about 4.8 times greater than the efficiency of one 7 cm3 detector. The resolution is 2.90 keV FWHM at 1.33 MeV and 2.01 keV FWHM at 122 keV. Some misalignment of the individual detector photopeaks is observed, probably caused by contact resistance. A technique which allows alignment of the peaks is presented.