DocumentCode :
788307
Title :
Characteristics of Coaxial Ge(Li) Detectors through Warm-Up and Re-Drift Cycles
Author :
Sakai, Eiji ; Fowler, I.L.
Author_Institution :
Physics Division, Chalk River Nuclear Laboratories, Atomic Energy of Canada Limited, Chalk River, Ontario, Canada
Volume :
15
Issue :
3
fYear :
1968
fDate :
6/1/1968 12:00:00 AM
Firstpage :
327
Lastpage :
331
Abstract :
Studies have been made of several cylindrical coaxial Ge(Li) detectors through various cycles of warming to room temperature and subsequent ´cleanup´ redrift - all in high vacuum. Conventional double-openend detectors with an n+-layer on the outside and a p-type core were used. Using the gamma-ray scanning technique it was observed that (1) When not fully cleaned up (capacitance decreasing with increasing voltage) the i-layer thickness increased from the n-layer towards the p-core as voltage was increased. (2) After a warm-up period, part of the i-layer nearest the core was lost with a reduction in efficiency (18% at 500V for one detector after an 8 hour warmup period). (3) Cleanup re-drift following a warm-up period restored the original characteristics but the cleanup time varied with the crystal from which the detector was made.
Keywords :
Coaxial components; Collimators; Energy resolution; Gamma ray detection; Gamma ray detectors; Geometry; Physics; Rivers; Temperature measurement; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1968.4324954
Filename :
4324954
Link To Document :
بازگشت