DocumentCode :
788450
Title :
Coincidence Experiments with Coaxial Ge(Li) Detectors in the Low Energy Range
Author :
Ostertag, E. ; Miehe, J.A. ; Henck, R. ; Siffert, P.
Author_Institution :
Centre de Recherches Nucléaires Département de Physique des Rayonnements et d´´Electronique Nucléaire Strasbourg - Cronenbourg France
Volume :
15
Issue :
3
fYear :
1968
fDate :
6/1/1968 12:00:00 AM
Firstpage :
413
Lastpage :
418
Abstract :
A ¿-¿ coincidence arrangement employing two large volume (20 cm3) Ge(Li) coaxial detectors has been realized. Each diode has two isolated outputs, one of which yields the energy information of the incident radiation and the other one the timing signal. The prompt resolution curve obtained with a 60Co source has a full width at half maximum of 11 ns. Coincidence ¿-ray spectra are observed with one Ge(Li) detector when the other is gated on the 1333 keV full energy peak. The use of a 30 ns coincidence resolving time results in an undistorted coincidence spectrum for energies higher than about 150 keV. In order to illustrate the applications of this technique, results obtained in ¿-¿ coincidence experiments with the complex ¿-ray emitter 132I are presented and discussed.
Keywords :
Circuits; Coaxial components; Detectors; Diodes; Electrodes; Energy resolution; Pulse amplifiers; Resistors; Signal resolution; Timing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1968.4324966
Filename :
4324966
Link To Document :