• DocumentCode
    789644
  • Title

    Detection of Low Energy Heavy Particles with Silicon Barrier Detectors

  • Author

    Ray, J.A. ; Barnett, C.F.

  • Author_Institution
    Oak Ridge National Laboratory Oak Ridge, Tennessee
  • Volume
    16
  • Issue
    1
  • fYear
    1969
  • Firstpage
    82
  • Lastpage
    86
  • Abstract
    The performance characteristics of silicon surface barrier detectors have been investigated for detecting H+, He+, N+, Ne+ and Ar+ in the energy interval 6-100 keV. For a cryogenically cooled, 50 mm2, 1325¿ depletion depth detector the resolution for protons was 1.65 keV (FWHM). The resolution decreased as the particle mass increased being 6.5 keV for 40 keV Ar+. All particles exhibit pulse height linearity throughout the energy range.
  • Keywords
    Argon; Atomic measurements; Detectors; Energy measurement; Energy resolution; FETs; Nitrogen; Pulse amplifiers; Silicon; Vacuum systems;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1969.4325084
  • Filename
    4325084